Used in medical institutions, in conjunction with the Electronic Endoscope Image Processor produced by our company, for imaging in the examination, diagnosis, or treatment of urinary tract diseases in the urinary system. Cannot be used with high-frequency accessories.
2. Product Information
Product Name: Disposable Ureterorenoscopes Catheter
Applicable Model: SU-ⅠR、SU-ⅠL、SU-ⅡR、SU-ⅡL、SU-ⅢR、SU-ⅢL、SU-ⅣR、SU-ⅣL
Product Specification Number: MPA.AH Certified No.20232060128
Registration Number: MPA.AH Certified No.20232060128
3. Product Performance
|
Models |
SU-ⅠR |
SU-ⅠL |
SU-ⅡR |
SU-ⅡL |
SU-ⅢR |
SU-ⅢL |
SU-ⅣR |
SU-ⅣL |
|
outer diameter of insertion tube outer diameter |
3.05mm |
2.85mm |
5.0mm |
|||||
|
Outer diameter of connection |
3.2mm |
3.0mm |
6.0mm |
5.4mm |
||||
|
Outside diameter of tip |
3.0mm |
2.95mm |
5.4mm |
4.4mm |
||||
|
Outside diameter of bending section |
3.05mm |
2.9mm |
6.0mm |
5.0mm |
||||
|
Instrument channel diameter |
1.13mm |
1.0mm |
2.3mm |
|||||
|
Working Length |
670 mm |
380 mm |
||||||
|
Bending angle |
Up:285° Down:285° |
Up:210° Down:210° |
||||||
|
Field of view |
110° |
|||||||
|
Depth of field |
2~50mm |
|||||||
|
Relative self-effect of illumination light |
At 90% of the field of view, the nominal value of the relative self-effect of illumination light luminosity is 0.4 |
|||||||
|
Luminance response characteristic |
The output signal of the electronic endoscope should maintain good linearity between the brightness calculated based on the electro-optic characteristics and the actual brightness of each gray scale of the measured board, and the linear fitting coefficient R2should not be less than 0.98. |
|||||||
|
Signal-to-noise ratio |
In the camera mode,the nominal values of signal-to-noise ratio is 35dB. |
|||||||
|
SFR(Spatial Frequency Respose) |
In the camera mode, the nominal values of SFR of 50% and 30% are 0.88C/° and 1.07C/°, respectively. |
|||||||
|
Static image tolerance |
In the camera mode,the nominal values of static image tolerance is 50. |
|||||||
|
Working condition |
Ambient temperature: 10℃~30℃; Relative humidity: ≤70% |
|||||||
|
Degree of protection against liquid ingress |
The insertion section is IPX7, and the rest is IPX0 |
|||||||
|
Degree of electric shock prevention |
After connected with the Electronic Endoscope Imaging Processor, it becomes the BF type application part. |
|||||||