1. Scope of Application
This product is inserted into the urinary tract or renal pelvis through the urethra or percutaneous insertion into the renal pelvis to observe, diagnose, photograph, and treat the urethra and renal pelvis. Cannot be used in conjunction with high-frequency diagnostic and therapeutic accessories.
Product Name: Flexible Video-Ureteroscope
Applicable Models: DU-IR、DU-IL、DU-ⅡR、DU-ⅡL
Product Specification Number: MPA.AH Certified No. 20222060016
Registration Number: MPA.AH Certified No. 20222060016
3. Product Performance
|
Models |
DU-IR |
DU-IL |
DU-ⅡR |
DU-ⅡL |
|
outer diameter of insertion tube outer diameter |
2.9mm |
2.7mm |
||
|
Outside diameter of tip |
3.0mm |
2.9mm |
||
|
Outside diameter of bending section |
3.3mm |
3.2mm |
||
|
Outer diameter of connection |
3.3mm |
3.2mm |
||
|
Instrument channel diameter |
1.16mm |
1.13mm |
||
|
Working Length |
700mm |
|||
|
Bending angle |
Up:276°,Down:276° |
|||
|
Field of view |
110° |
|||
|
Depth of field |
2~50mm |
|||
|
Relative self-effect of illumination light luminosity |
At 90% of the field of view, the nominal value of the relative self-effect of illumination light luminosity is 0.215 |
|||
|
Luminance response characteristic |
The output signal of the electronic endoscope should maintain good linearity between the brightness calculated based on the electro-optic characteristics and the actual brightness of each gray scale of the measured board, and the linear fitting coefficient R2should not be less than 0.98. |
|||
|
Signal-to-noise ratio |
In the camera mode,the nominal values of signal-to-noise ratio is 35dB. |
|||
|
SFR(Spatial Frequency Respose) |
In the camera mode, the nominal values of SFR of 50% and 30% are 0.88C/° and 1.07C/°, respectively. |
|||
|
Static image tolerance |
In the camera mode,the nominal values of static image tolerance is 50. |
|||
|
Degree of protection against liquid ingress |
IPX7 |
|||
|
Working condition |
Ambient temperature: 10℃~30℃; Relative humidity: ≤70% |
|||